PRODUCTS
sterillity testing isolator cabinet
Description
sterillity testing isolator cabinet.
STl series sterile isolator is a controlled clean micro-environment that not only removes the external microbial contamination,isolation,and non-personnel contamination,but also maintains the cleanliness level A environment. It is equipped with a special hydrogen peroxide VHP sterilizer,a residue removal system after sterilization,a sterile air supply system,a closed cabin system,a special HMl operation and recording system,and a computerized verification system software. The sterile isolator is widely used in pharmaceutical,medical Sterile isolation operation in sanitary,biological test and other places.
Automatic leak detection:inflate the cabin and check whether the system is closed. If the airtight requirements are met,the sterilization process will be started. If the airtight requirements are not met,continue to inflate and seal,and then leak detection,and so on,until the airtight conditions are met.
High-efficiency flash evaporation method: high-precision constant temperature system ensures that hydrogen peroxide will decompose at 0% and form steam at 100% during evaporation.
Rapid residue removal: the configured hydrogen peroxide decomposer can not only rapidly and completely decompose the hydrogen peroxide in space, but also decompose the hydrogen peroxide residue on the surface.
sterilization process record: automaticaly detect sterilization concentration and automatically draw sterilization process curve.
Door interlock function: the door cannot be opened during the operation of the equipment.
Bacteriostatic treatment of the interior surface of the cabin: the interior surface of the cabin is treated with bacteriostasis, which can achieve sterilization and bacteriostasis, so that the interior surface of the cabin will never have a place for microorganisms to hide.
Conform to computerized verification software:the system has the functions of audit trail,electronic signature and data integrity.
Features of bottom-mounted transmission cabin:the bottom-mounted transmission cabin has compact structure,small floor area and easy operation.
TECHNICAL CHARACTERISTICS
material | All 304 stainless steel outside and 316L stainless steel inside |
---|---|
cleanliness level | CLASS A |
wind speed | 0.45m/s±20% |
Leakage rate | ≤0.5% VOL/h(below 100pa) |
Exhaust volume | ≥300m³/h |
Pressure control range | -100pa~+100pa |
Sterilization efficiency | 6lg |
VHP residue removal efficiency | ≤1PPM |
control mode | Manual or automatic |
noise | ≤68dB(A) |
illumination | ≥350LUX |
model / parameter | Overall dimension (W*D*H)mm | Operation cabin size(W*D*H)mm | Size of transfer cabin(W*D*H)mm | power supply | Compressed air volume | explain |
---|---|---|---|---|---|---|
ST1-1200 | 1600*880*2250 | 1195*680*700 | AC220V/3.5KW | ≥0.6Mpa | Single cabin | |
STI-1200/BPB600 | 1600*880*2250 | 600*680*700 | Single and double cabin | |||
ST1-1500D/BPB600 | 2100*1200*2250 | 1495*1195*700 | 600*680*700 | Single and double cabin | ||
ST1-1800 | 2200*880*2250 | 1795*680*700 |
AC220V/5.5KW
| Double person and single cabin | ||
STI-1800/BPB600 | 2200*880*2250 |
1795*680*700
| 600*680*700 | Double person and single cabin | ||
ST1-1800/PB600 | 2800*880*2250 | 600*680*700 | Double cabin | |||
ST1-1800D/BPB600 | 2200*1200*2250 | 600*680*700 | Two people,two cabins, two sides |