VHP series vaporized hydrogen peroxide sterilizer

VHPS sterilizer is an equipment that forms hydrogen peroxide steam from H202 solution (i.e.hydrogen peroxide) through vaporizer, and then transmits it to the space to be sterilized through pipeline for biological decontamination, lt is widely used for room temperature sterilization otconfined space, including transfer tank, filling machine, freeze-drying machine, isolator, clean room and other places.

Description

Patent:
ZL201610220343.8
ZL201610392506.0
ZL201320372742.8
ZL201620538698.7

 

Product introduction.

VHPS sterilizer is an equipment that forms hydrogen peroxide steam from H202 solution (i.e.hydrogen peroxide) through vaporizer, and then transmits it to the space to be sterilized through pipeline for biological decontamination, lt is widely used for room temperature sterilization otconfined space, including transfer tank, filling machine, freeze-drying machine, isolator, clean room and other places.

Brief description of working process.
Biological decontamination process of hydrogen peroxide sterilizer:
● system preheating - connect the inlet and outlet pipes of the sterilizer to the inlet and outlet interfaces of the closed cabin, start the heater, and conduct preheating and dehumidification treatment in the sterilization cabin;
● formation of Vaporized hydrogen peroxide aerosol - start the sterilization procedure, input the hydrogen peroxide solution into the vaporizer to form submicron particles of hydrogen peroxide, and form hydrogen peroxide aerosol, which is input into the closed space to be sterilized by the anti-corrosion fan for 60-90 minutes;
● residue removal stage - start the residue removal procedure, switch the air flow to the hydrogen peroxide residue remover, the hydrogen peroxide in the sterilization chamber is decomposed into water and oxygen, and the hydrogen peroxide concentration decreases rapidly.
When the concentration drops below 1ppm, the system stops running and the sterilization chamber can be opened.

 

Performance parameter.

Model / parameter  

VHPS-50

VHPS-100

VHPS-200 

          VHPS-500                                                

Biological decontamination ability                                          

6Lg

Rated vaporization capacity

0-5g/min

0-10g/min

0-20g/min

                          0-30g/min                       

Sterilization factor Hydrogen peroxide vapor
Maximum sterilization space volume

60m³/h

120m³/h

200~250m³/h 

 700m³/h

Overall dimension (mm)

600*720*1150(W*D*H)

740*870*1150(W*D*H)

   1200*900*1950(W*D*H)    

1400*1150*1950(W*D*H)   
Weight (kg)

150 

200

300 

  400
Power                  2.5KW                 12KW                     8KW   35KW
Power supply 

AC220V/50Hz

 

 

 

 

 

 

 

 

 

 

 

 

 

Technical features.
1. Four unique technologies and features:
 unique hydrogen peroxide dry fog generation technology: convert hydrogen peroxide liquid into hydrogen peroxide vapor aerosol through vaporizer at room temperature and suspend it in the air.
 efficient biological decontamination capacity: hydrogen peroxide vapor aerosol can quickly and completely oxidize microorganisms on the surface into water, oxygen and carbon dioxide, with short sterilization time and high efficiency.
 it can be equipped with a residue remover with unique residue removal function: the hydrogen peroxide remover can quickly reduce the hydrogen peroxide Vaporized to water and oxygen, with short residue removal time, high efficiency and no risk.
2. Functional features:
 wide sterilization range, which can quickly sterilize large confined spaces;
 low temperature sterilization process, room temperature is enough;
 equipped with residue remover, which can decompose the final product of hydrogen peroxide into water vapor and oxygen;
 short sterilization time and low operation cost;
 start the machine regularly and set the sterilization time;
 it can detect the hydrogen peroxide concentration in the sterilization process in real time and draw the hydrogen peroxide concentration curve.
 
Use environment
● ambient temperature: 10 ℃ ~ 40 ℃;
● relative humidity: ≤ 85% RH;
● voltage AC220V±10%; Frequency 50Hz ± 1Hz; Effective grounding;
● relatively sealed environment;
● no dust and smoke environment,