PRODUCTS


+
  • Y09-SS(1726233450986).jpg
  • Y09-SS(1726233451049).jpg
  • Y09-SS(1726233451084).jpg

Y09-SS

Y09-SS type/5102 online laser dust particle counter is used to measure the size and number of dust particles per unit volume of air in a clean environment, and transmit the corresponding particle number to the host computer through MODBUS485 communication. It can control the clean environment with cleanliness level of 100,000 to 100. This instrument uses imported semiconductor laser light source, small size, light weight, high detection accuracy, computer control, as long as the instrument is powered on, it can automatically output data without personnel operation. Therefore, it is very convenient to test a clean environment. It is widely used in electronics, optics, chemistry, food, cosmetics, medicine and health, biological products, aerospace and other sectors.

Description

Introduction to Counter Application

Y09-SS type/5102 online laser dust particle counter is used to measure the size and number of dust particles per unit volume of air in a clean environment, and transmit the corresponding particle number to the host computer through MODBUS485 communication. It can control the clean environment with cleanliness level of 100,000 to 100.
This instrument uses imported semiconductor laser light source, small size, light weight, high detection accuracy, computer control, as long as the instrument is powered on, it can automatically output data without personnel operation. Therefore, it is very convenient to test a clean environment. It is widely used in electronics, optics, chemistry, food, cosmetics, medicine and health, biological products, aerospace and other sectors.
 
 
The main technical parameters
 
1. Dimensions      ø80*160mm; 140*100*45
2. Mass        about 0.6kg
3. Maximum power consumption     2VA
4. Power supply     external 7.4-12VDC/1A
5. Standard signal particle size channel  0.3-5μm optional 2, standard 0.5, 5μm
  6. Sampling flow   28.3LPM
  7. Sampling period 1 minute
  8. Use environmental conditions   Temperature: 0℃-40℃          Humidity: 20%-75% ambient pressure and sampled gas pressure: 86kPa-106kPa
9. The maximum allowable sampling concentration is  35000 particles/L (the particle size of dust particles is not greater than 0.5μm), and the sampling air must not contain corrosive gases such as acid and alkali
10. Self-cleaning time     ≤10min
 
 
working principle
 
This instrument uses the principle of light scattering. When suspended particles in the air pass through the photosensitive area, they scatter light flux in proportion to their particle size. After photoelectric conversion, amplification and processing, the equivalent diameter and number of collected particles are obtained and the number of particles is passed through MODBUS485 transmission
 
 
Install
Know the counter before installation, as shown in the figure below. It is composed of protective cover, isokinetic sampling port, entity, electrical connection plug, sampling hose, ø32 stainless steel tube, quick connector, etc.