Fume Hood

It is widely used in scientific research, biological & chemical experiment, electron, precise apparatus, instrument and so on. To Control exhausting the organic solvent and gas with poisonous、deleterious、flammable, to protect operator .

Description

Product introduction

It is widely used in scientific research, biological & chemical
experiment, electron, precise apparatus, instrument and so on.
To Control exhausting the organic solvent and gas with
poisonous、deleterious、flammable, to protect operator .

 

Performance Parameter

Parameter\Model TGH-1200 TGH-1500 TGH-1800
Air displacement 1200m³ /h  1500m³ /h  2000m³ /h 
Air tightness 500Pa leakage rate≤10%(in 30min)
Equiblast air volume ≥0.4m/s
Noise ≤62dB(A)
Half peak value of shaking of shaking ≤4μm
Illumination ≥300Lx
Power Supply AC220V50Hz
Max. power consumption 0.4KW 0.6KW 0.8KW
Weight 150kg 200kg 250kg
Dimension of working area 1000*650*800 1300*650*800 1600*650*800
Overall dimension 1200*800*2350 1500*800*2350 1800*800*2350
Fluorescent lamp 30W*2 40W*2 40W*3